Operation
Doc. 031880-04 1/12
35
In the example program shown in Figure 39, TTL 1 output on the
pump (which is connected to the Dionex RFC
Gradient
TTL input) is
turned on at 2.00 minutes and turned off at 14.00 minutes.
Figure 39. PeakNet 6 Example Program for Gradient Control
Front Panel Method Gradient Example
NOTE Front panel control is not available for the DX-120 TTL
outputs.
This example shows how to edit or create a front panel Method that
includes timed events for controlling the Dionex RFC
Gradient
TTL
input. In this example, the Dionex RFC Gradient TTL input is connected
to the TTL1 output on an IS25 pump.
1. From the front panel of the pump, go to the
METHOD
screen.
2. Open a Method that you want to edit, or start a new Method.
NOTE For a gradient pump or combination pump/detector
module, press the right arrow button to display the
extension to the
METHOD
screen, which includes the
fields for controlling TTL and Relay outputs.
Turns on the Gradient TTL
Turns off the Gradient TTL