2 • DC Description
Doc. 065342-06 12/11
75
•
The eluent/sample mixture is pumped through a temperature stabilizer
(analytical IC only), through the guard and separator columns
, and
then through the suppressor
.
•
From the suppressor, the mixture flows through a CRD
(if installed) to
the cell
, where the analytes are detected. A digital signal is sent to
Chromeleon software.
•
The mixture flows out of the cell and is recycled back into the
suppressor
, where it is used as the water source for the regenerant
chamber. Regenerant flow is routed through the CRD
(if installed),
through the EG degas tubing
, through the CR-TC
, and then out to
waste
.
2.15
ED Electrochemical Detector
One or two Dionex ICS-5000 Electrochemical Detectors (EDs) can be
installed in the DC. Each complete ED assembly consists of an amperometric
detection cell and the detector electronics required to collect data and send it
to the computer and the analog output (if installed). The detector is installed in
the upper compartment of the DC (see
). The cell mounts on the
front of the detector. The detector electronics are not accessible to the user.
The ED can perform the following electrochemical detection modes:
•
DC amperometry (see
•
Integrated amperometry—includes pulsed amperometric detection (PAD)
and integrated pulsed amperometric detection (IPAD) (see
•
2.15.1 Electrochemical Detector Cell
The ED cell includes three different electrodes. It is a miniature flow-through
amperometric detection cell with a titanium cell body (the counter electrode),
a working electrode, and a combination pH-Ag/AgCl reference electrode (see
) or a PdH reference electrode. The type of working electrode
used depends on the application. Four conventional (non-disposable) working
electrode types are available: gold, platinum, silver, and glassy carbon. Five
disposable working electrode types are available: gold on two different
substrates (polyester or PTFE), silver, platinum, and carbon.
6
7
8
9
10
11
12
13
14
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...