2 • Description
Doc. 065342-06 12/11
31
NOTE
For routine on/off control, use the
POWER
button on the front
of the DP/SP (see
and
). To turn off the
pump, press and hold the
POWER
button for 2 seconds.
The fuse cartridge contains two 2-amp IEC 60127-2 slow-blow fuses
(P/N 954773). For instructions on how to change the fuses, see
.
The power cord plugs into the IEC 320 three-prong receptacle.
Digital I/O Port
The digital I/O port provides a connection to auxiliary devices. The connector
includes three TTL inputs and four relay outputs.
indicates the
functions assigned to the connector pins.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the DC and is easily accessible.
Le cordon d'alimentation principal est utilisé comme dispositif principal
de débranchement. Veillez à ce que la prise de base soit située/installée
près du module et facilement accessible.
Das Netzkabel ist das wichtigste Mittel zur Stromunterbrechung. Stellen
Sie sicher, daß sich die Steckdose nahe am Gerät befindet und leicht
zugänglich ist.
The maximum switching voltage of the relays is 24 V. The switching
current must not exceed 100 mA.
Pin Number
Signal Name
Signal Level
Description
1
-----
-----
Not used
2
-----
-----
Not used
3
Relay 3 Out
Potential-free
Normally open
4
Relay 1 Out
Potential-free
Normally closed
5
Relay 2 Out
Potential-free
Normally closed
6
Relay 3 Out
Potential-free
Normally closed
7
Relay 1 Out
Potential-free
Common
Table 2-1. 25-Pin D-Sub I/O Port (Female)
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...