5 • Operation
Doc. 065342-06 12/11
179
to the sample port on the injection valve. If not, refer to
for
connection instructions.
2. Fill a syringe with a calibration standard or sample.
3. Insert the syringe into the sample loading port on the front of the DC.
4. Verify that the injection valve is in the Load position. If it is not, switch
the valve by clicking the valve control on the Chromeleon 7 ePanel Set or
the Chromeleon 6.8 panel tabset, or by pressing the
VALVE
button on the
front of the DC.
5. Overfill the sample loop with several sample loop volumes. Excess
sample will exit through the injection valve waste line.
6. Leave the syringe in the port. This prevents the sample from exiting the
loop before injection.
7. Switch the injection valve to the Inject position.
5.3.3
Loading Samples with a Vacuum Syringe (Pull Method)
1. Verify that a luer adapter fitting is installed on the sample loading port on
the front of the DC and that tubing is connecting the sample loading port
and the injection valve. If this is not the case, refer to
connection instructions.
2. Disconnect the waste line from the injection valve and replace it with a 25
to 30 cm (10 to 12 in) piece of PEEK or PTFE tubing (see
3. Place the free end of the line into the sample.
4. Verify that the injection valve is in the Load position. If it is not, switch
the valve by clicking the valve control on the Chromeleon 7 ePanel Set or
the Chromeleon 6.8 panel tabset, or by pressing the
VALVE
button on the
front of the DC.
5. Insert a 5 cc syringe into the sample loading port on the DC front door
) and pull out the plunger to draw the sample into the
injection valve.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...