Dionex ICS-5000
+
Installation Instructions
46
Doc. 065447-02 12/14
•
The DP, EG, SP, and TC Ship Kits include a 1 m (3 ft) USB cable
(P/N 063246).
3.8.2
USB Compatibility
All Dionex ICS-5000
+
modules, as well as other Dionex modules connected
to them (for example, an autosampler), are equipped with USB receptacles.
Your system may include a combination of the following:
•
Dionex modules with an internal USB 2.0 hub (powered)
•
Dionex modules with an internal USB 1.1 hub (unpowered)
•
Dionex modules with no internal USB hub
•
An external USB 2.0 hub
To avoid problems caused by incompatibility between USB 2.0 and USB 1.1,
identify the USB version of each system component to be installed (see
and observe the rules for connections shown in
.
The USB standard limits the USB cable length to 5 m (5.5 yds). Each
USB module can be separated from the PC by no more than five hubs.
Thus, each USB module can be located no more than 30 m (32 yds)
from the PC.
Type of Hub
Module
Notes
USB 2.0 internal
hub (powered)
Dionex ICS-5000
+
DC
Dionex ICS-5000
+
EG
When you turn off the power using the
POWER
button on the front of the module
(and the main power cord is connected), the
USB 2.0 hub remains powered on. Thus, all
modules connected downstream will remain
connected to Chromeleon.
USB 1.1 internal
hub
(unpowered)
Dionex AS-AP
Dionex ICS-5000
+
DP
Dionex ICS-5000
+
SP
When you turn off the power using the
POWER
button on the front of the module,
the USB 1.1 hub is powered off and all
modules downstream are disconnected
from Chromeleon.
No internal hub
Dionex ICS-5000
+
TC
Dionex PDA
Dionex VWD
Table 3-2. USB Versions of Dionex Modules