Dionex ICS-5000
+
Installation Instructions
98
Doc. 065447-02 12/14
4. For a Dionex ATC 500, ATC-HC 500, ATC-HC Borate, or CTC 500:
a.
Locate the mounting clip required for your column in the pump ship
kit.
b. Remove the screw and mounting clip used to secure the static mixer
to the component panel. Attach the mounting clip for the trap column
to the component panel, using the screw from the mixer clip.
NOTE
When using a dual-component clip, you can install the
trap column in either set of clips.
c.
Connect the tubing from the secondary pump head to the trap column
inlet and press the trap column into the mounting clip.
5. Cut an 86 cm (34 in) piece of 0.25 mm (0.010 in) ID black PEEK tubing
(for standard bore systems) or 0.125 mm (0.005 in) ID red PEEK tubing
(for microbore systems).
6. Connect the tubing to the trap column outlet.
7. Store the static mixer (and its mounting clip, if removed) for future use.
4.2.4
Completing the IQ Checklist
Complete the Installation Qualification (IQ) checklist to confirm that the
Dionex ICS-5000
+
system is correctly installed. For instructions, refer to
Dionex ICS-5000
+
System Installation Qualification
(Document
No. 065402), provided in the DC Ship Kit (P/N 072011, standard DC;
P/N 22171-62000, low-temperature DC).
Thermo Scientific High-Pressure Trap
Column
Use This Mounting Clip ...
Dionex ATC 500, 2 mm (P/N 079018)
Dionex CTC 500, 2 mm (P/N 079019)
Dual-component clip
(P/N 074010)
Dionex ATC 500, 4 mm (P/N 075976)
Dionex CTC 500, 4 mm (P/N 075977)
Component clip
(P/N 075985)
Dionex ATC-HC 500 (P/N 075978)
Dionex ATC-HC 500 Borate (P/N 075979)
Component clip
(P/N 075985)