3 • Operation and Maintenance
Doc. 031857-03 4/06
3-17
•
Enter the commands for loading the sample on the autosampler
front panel.
4. Switch the injection valve to the Inject position (see
3.10.4 Injecting Samples
After loading the sample in the sample loop, use one of the following
methods to switch the injection valve to the Inject position:
•
Manually: Click the
Inject
button on the Chromeleon Control panel
) or touch the
Inject
button on the
HOME
page.
•
Automatically: Include the Inject command in a Chromeleon
program. See the examples in
.
3.10.5 Example Chromeleon Commands for Loading and
Injecting Samples
The following examples show commands for loading and injecting
samples using an AS or AS or AS50 (USB)40 autosampler.
Example AS or AS50 (USB) Program Commands
0.000
Pump_ECD.Autozero
;Zero the baseline.
Load
;Switch the valve to Load.
Wait
CycleTimeState
;Wait for cycle time (if any).
Inject
;Switch the valve to Inject.
Wait
InjectState
;Wait for injection to complete.
ECD_1.AcqOn
;Start data acquisition.
14.000
ECD_1.AcqOff
;Stop data acquisition.