C • Touch Screen Operation
Doc. 031857-03 4/06
C-11
C.5
Plot Page
The
PLOT
page provides a real-time plot of the detector conductivity output. To
open the page, touch the
PLOT
button on the
HOME
page, or select
PLOT
from the
menu of pages. The page can display up to 60 minutes of data. Data older than
60 minutes is discarded. You can view the plot and adjust the display when the
ICS-2000 is under either local or computer control.
Plot Controls
•
The
plot offset
button offsets the current background conductivity reading
and sets the plot baseline to zero. Subsequent output is plotted in relation to
the new zero point. Offsetting the plot does not affect the data sent to
Chromeleon, the analog output, or the data displayed on other pages.
•
The
range
field sets the highest conductivity reading displayed on the plot
(the vertical scale). If the tops of peaks are cut off, select a higher range. If
peaks are short, select a lower range. Select
Auto
to have the scaling of the
signal plot automatically adjusted to the height of the signal.
•
The
elapsed time
field displays the time since injection.
•
The
plot time
field sets the length of time plotted on the page (the horizontal
scale).
Figure C-8. Plot Page
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...