Connecting the Waste Lines
Doc. 031857-03 4/06
B-27
B.10 Connecting the Waste Lines
The ICS-2000 has the following waste lines:
•
PUMP WASTE
carries prime waste from the pump.
•
TO WASTE OUT
carries sample overflow from the injection valve.
•
WASTE, GAS SEPARATOR
carries system waste from the suppressor.
1. Untape the coiled waste lines from the rear panel. Place the ends of the
PUMP
WASTE
and
TO WASTE OUT
lines into a waste container.
2. Direct the
WASTE, GAS SEPARATOR
line from the top of the ICS-2000 to the
rear panel. Snap the line onto one of the tubing clips on the rear panel.
Connect the line to the gas separator waste tube (see
NOTE To prevent waste from siphoning back into the system,
check the lines periodically to be sure they are not bent,
pinched, or elevated at any point.
B.10.1 Installing the Gas Separator Waste Tube
1. Locate the gas
separator waste tube
assembly in the ICS-
2000 Ship Kit
(P/N 064375).
2. Connect the
WASTE,
GAS SEPARATOR
line from the
suppressor to one of
the 3-mm (1/8-in) ID
white Teflon lines on
the gas separator
waste tube assembly
(see
Figure B-24. Gas Separator Waste Tube Installation
3. Place the gas separator waste tube assembly into the waste container.
Make sure the tubing junction (where the white Teflon tubing meets
the black polyethylene tubing) is above the top of the container and
Gas Separator
Waste Tube
Assembly
Waste
Waste, Gas
Separator Line
(from the suppressor)
This line is
provided for a
second system
if needed.
Open to
the air
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...