ICS-2000 Ion Chromatography System
5-70
Doc. 031857-03 4/06
4. Connect the
TO EGC CO3 MIXER IN
line to the mixer inlet and connect
the
TO EGC CO3 MIXER OUT
line to the mixer outlet.
5.20.2 Filling the EGC CO
3
Mixer with Deionized Water
Before initial use, follow the steps below to temporarily fill the EGC CO
3
Mixer with deionized water. Then, follow the steps in the next section to
fill the mixer with eluent of the required concentration for the application
to be run.
1. Verify that the EGC-1 and EGC-2 currents are off.
2. Disconnect the EluGen cartridge inlet line (labeled
TO
PUMP/DAMPER
) from the coupler that connects it to the green pump
Figure 5-62. ICS-2000 with an EGC CO
3
Mixer Installed
TO EGC CO3
MIXER IN
TO EGC CO3
MIXER OUT