5 • Service
Doc. 031857-03 4/06
5-35
5.14 Replacing the Leak Sensor
1. Turn off the power to the ICS-
2000.
2. Open the ICS-2000 front door.
3. Loosen the screw on the front of
the leak sensor (see
Note:
The screw
remains attached to the sensor.
Figure 5-27. Leak Sensor
4. Remove the leak sensor from
the component panel and pull
the cable out of the opening in
the panel.
5. The cable attached to the leak
sensor is connected to a
matching cable inside the ICS-
2000. Continue pulling the cable
until the connectors for the two
cables are outside of the panel
(see
6. Disconnect the two cables.
Figure 5-28. Leak Sensor Cable
7. Connect the cable from the new leak sensor (P/N 058053) to the ICS-2000
cable.
8. Feed the cables back inside the ICS-2000. Align the leak sensor with the
component panel opening and fingertighten the screw.
9. Make sure the leak sensor does not touch the bottom of the drip tray.
Leak
Sensor
Loosen
Screw
Disconnect
Cables