3 • Operation and Maintenance
Doc. 065330-05 03/2020
57
3.6.3
Stopping a Run
To stop a run before the end of the method, press the
PAUSE
and
CONT
buttons at the same time. The system will start the abort process without
further warning and display a screen with two options:
•
To immediately return the Dionex AutoTrace 280 to an idle state,
press
PAUSE
.
•
To rinse all sample lines and clear solvent lines of any remaining
solvent before returning the Dionex AutoTrace 280 to an idle state,
press
CONT
.
3.7
Post-Run Activities
After running a method:
•
Make any necessary procedural changes (during method development)
•
Test the changes
•
Remove the used cartridges or disks
•
Prepare to run the next batch of samples
•
Clean up, if done for the day
3.7.1
Determining the Required Changes
Based on the recovery data and operation observations during method
development, you may need to modify the Dionex AutoTrace 280
method. The
Elute Flow
and
Load Flow
parameters are usually the most
critical settings.
For additional assistance, refer to the following sources: the method
development guide published by the manufacturer of your SPE cartridges
or disks, the application tips in
, and the AutoTrace software
Help.
NOTE When attempting to optimize a method, making
changes one at a time is the best way to evaluate the
effect of a change.