Dionex AS-DV Operator’s Manual
82
Doc. 065259-05 10/12
5. Remove the 6-port or 10-port high-pressure valve:
a. Use a 3-mm Allen wrench to remove the two screws holding the valve
assembly in place (see
b. Slide the valve assembly partway out of the chassis, press the locking
latch on the cable (see
), and disconnect the valve cable.
c. Remove the valve assembly.
6. Unscrew the black locking ring on the front of the valve (see
) and
remove the ring.
Figure 5-15. Removing the Valve Assembly
Figure 5-16. Disconnecting the Valve Cable
Remove these
2 screws
Black
Locking Ring
Locking Latch