This error occurs if there is a discrepancy between the vial size installed in the
carousel and the vial size currently stored in Dionex AS-DV memory. (For
example, a 5.0 mL vial is installed in position 10, but the Dionex AS-DV
expects a 0.5 mL vial to be installed in that position.) Vial memory must be
reset before starting a new run.
To troubleshoot:
1. Before starting a new run, press the
Carousel Release
button to reset the
vial memory.
2. If you want to replace vials during a run, use the controls on the Dionex
AS-DV ePanel (in Chromeleon 7) or Control panel (in Chromeleon 6.8)
under
Reset Vial Memory
to reset the memory for only the vials you
want to replace. For instructions, see
3. If the problem persists, contact Technical Support for Dionex products for
assistance.
4.2
Liquid Leaks
•
Old, previously-used caps or sample vials
Do not reuse caps or sample vials.
•
Damaged tubing between sampling tip and optional high-pressure valve
in the Dionex AS-DV or injection valve in the IC system
Replace the sampling tip and tubing (see
The vial size has been changed without resetting the
carousel.
Before starting a run, always press the
Carousel Release
button to
disengage the carousel and reset vial memory. After loading the vials
in the carousel, press the
Carousel Release
button again to engage
the carousel.
During a run, do not press the Carousel Release button. Doing so
clears all current vial information from Dionex AS-DV memory. During
a run, the Dionex AS-DV stores the vial size installed in each carousel
position and the remaining volume in each vial.
Summary of Contents for Dionex AS-DV
Page 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Page 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Page 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Page 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...