Example: Using the From Sequence Volume Option
When
From Sequence
is selected for the sample delivery volume, the
Dionex AS-DV calculates the delivery volume using the formula shown
in the
Total Volume Delivered Estimator
. In this case, the
DeliverSample
command is included without a specified volume.
In the example shown in
, the loop volume is 25
P
L, the delay
volume is 125
P
L (the default), and the flush factor is 8. This provides a
delivery volume of 1200
P
L. This example allows four injections per
filled 5.0 mL vial.
Notes:
•
In
Loop Mode
, you can increase or decrease the delivery volume by
entering a different
Flush Factor
.
Figure 3-17. Sample Preparation Options (Chromeleon 6.8 Version Shown):
Estimating the Delivery Volume Example