Installation
The Gauge Head
The Gauge Head
Warning
Installation must be in accordance with local and national
electric codes for the area classifications.
▲
Warning
The source housing handle must be in the OFF position during
installation.
▲
Warning
Do not reach inside the source housing at any time during
installation.
▲
Warning
Use proper lifting procedures during installation to avoid
injury.
▲
Mounting
Configurations
unting
Configurations
The optimum gauge mounting configuration depends on the application;
however, the source head and the detector are typically mounted together
on the opposite sides of a pipe. The following mounting options are
supported for the gauge:
●
Dual chain
●
Pipe saddle (cradle) mount
●
Pipe spool with gauge head pre-installed
The chain mount is the most frequently used mounting configuration, as it
allows the unit to be mounted on a range of pipe sizes using the same
mounting hardware.
Note
Moving the gauge to a pipe with a different diameter may require
changing the radiation source size and/or reconfiguring the detector (ion
chamber) electronics. Contact Thermo Fisher for assistance.
▲
In addition to the
listed earlier in this chapter, note the
following when planning gauge head installation.
●
Whichever mounting configuration is used, it is important to align the
beam path (the centerline of the source housing) as closely as possible
with the centerline of the detector housing. Be sure to mount the gauge
head securely, since any movement or change in alignment can affect
the gauge’s calibration.
Thermo Fisher Scientific
Den Installation Guide
3-7
Summary of Contents for DensityPRO+
Page 2: ......
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 14: ...This page intentionally left blank ...
Page 46: ...This page intentionally left blank ...
Page 48: ...This page intentionally left blank ...
Page 100: ...This page intentionally left blank ...