background image

Cimarec

®

Ceramic Top

Stirring Hot Plates,

Stirrers, and Hot Plates

OPERATION MANUAL

AND PARTS LIST

LTM1950X1 • 10/29/09

4” x 4” (10cm x 10cm)

HP194515 Hot Plate

S194615 Stirrer

SP194715 Hot Plate Stirrer

7” x 7” (18cm x 18cm)

HP194825 Hot Plate

S194925 Stirrer

SP195025 Hot Plate Stirrer

Summary of Contents for Cimarec HP194515

Page 1: ...tes Stirrers and Hot Plates OPERATION MANUAL AND PARTS LIST LTM1950X1 10 29 09 4 x 4 10cm x 10cm HP194515 Hot Plate S194615 Stirrer SP194715 Hot Plate Stirrer 7 x 7 18cm x 18cm HP194825 Hot Plate S194925 Stirrer SP195025 Hot Plate Stirrer ...

Page 2: ...ings 3 Operation 6 Introduction 6 Setting Heating Temperature 6 Setting Stirring Speed 6 Specifications 7 4 x 4 Models 7 7 x 7 Models 8 Replacement Parts Listings 9 Hot Plates 9 Stirrers 9 Stirring Hot Plates 9 Ordering Information 10 Warranty 11 ...

Page 3: ...re occurs the top surface tempera ture could rise to the maximum temperature 300 540 C depending on your model s specification and remain at that temperature indefinitely Under these conditions the material being heated on the surface of the hotplate could reach levels in excess of the maximum temperature Warnings Please note the following WARNINGS 1 Do not immerse unit for cleaning 2 As with all ...

Page 4: ...dry cloth 8 Replace the top immediately if damaged by etching scratching or chipping A damaged top can break in use 9 Do not use metal foil on hot plate Do not use metal containers sand bath or other insulating material on the hot plate or stirring hot plate Top plate can be damaged and shock hazard can exist 10 Do not use in highly corrosive atmospheres corrosive fumes and spill may damage top an...

Page 5: ...ust or respirable fibers resulting from operation of this equipment under normal con ditions However there may be a risk of exposure to res pirable dusts or fibers when repairing or maintaining the insulating materials or when otherwise disturbing the mate rials in a manner which causes release of dust or fibers therefrom Through the use of proper handling procedures you can work safely with these...

Page 6: ...se could cause damage to the hot plate and stirring hot plate Both the temperature and speed controls are combination on off switches and setting selectors Setting Heating Temperature To select a temperature turn the control clockwise to the desired setting The lighted heat indicator shows when the unit is operating Setting Stirring Speed To select a speed turn the speed control clockwise until th...

Page 7: ...75 2 2 Electrical Ratings Volts 120 120 120 Amps 3 21 0 45 3 45 Watts 385 21 405 Frequency 60 Hz 60 Hz 60 Hz Phase 1 1 1 Maximum operating temperature C F 538 1000 N A 538 1000 Maximum operating speed N A 2500 2500 Environmental Conditions Ambient temperature 17 C to 27 C 62 6 F to 80 6 F Relative humidity non condensing 20 to 80 Installation category Category II per IEC664 Pollution degree Degree...

Page 8: ...6 Electrical Ratings Volts 120 120 120 Amps 8 96 0 48 9 2 Watts 1062 23 1086 Frequency 60 Hz 60 Hz 60 Hz Phase 1 1 1 Maximum operating temperature C F 538 1000 N A 538 1000 Maximum operating speed N A 2500 2500 Environmental Conditions Ambient temperature 17 C to 27 C 62 6 F to 80 6 F Relative humidity non condensing 20 to 80 Installation category Category II per IEC664 Pollution degree Degree 2 p...

Page 9: ... Part Number 7x7 Part Number Description 540 0033 710 0017 Top plate only CN1928X1 CN1931X1 Control speed PLX108 PLX108 Pilot light CR1928X1 CR1928X1 Cord set 120 volt Stirring Hot Plates 4x4 Part Number 7x7 Part Number Description 540 0033 710 0017 Top plate only ELX71 ELX74 Element CNX171 CNX171 Control heat CN1928X1 CN1931X1 Control speed PLX108 PLX108 Pilot light CR1928X1 CR1928X1 Cord set 120...

Page 10: ...c dealer from whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request then contact our Customer Service Department at 563 556 2241 or 800 553 0039 Prior to returning any materials please contact our Customer Service Department for a Return Materials A...

Page 11: ... misuse including use inconsistent with written operating instructions for the product mishandling contamination overheating modification or alteration of the product by any customer or third party or iii use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific prod ucts Heating elements because of their susceptibility to overheating and contamin...

Reviews: