1100°C Box Furnace
330985H01 |
9-57
Maintenance |
Chapter 9
9. Replace the gas inlet tube assembly.
Figure 6
Thermocouple Replacement
Solid-State Relay Replacement
If the SSR is inoperable, complete the following steps to replace the relay (refer to
Figure 7 “Solid Relay Replacement”):
1. Remove the screws located on the left and right sides of the control panel (item #1 in
Figure 7).
2. Slide the panel assembly away from the unit to expose components.
3. Locate the SSR on the component tray (item #2 in Figure 7).
4. Note the terminal connections of the relay wires and label them for reattachment.
Remove the wires from the terminals of the relay.
5. Remove the mounting screws from the relay.
6. Replace the relay and reconnect the wires.
Summary of Contents for BF51728
Page 11: ...Safety Notes Chapter 1 1 10 330985H01 1100 C Box Furnace...
Page 15: ...Introduction 2 14 330985H01 1100 C Box Furnace Chapter 2...
Page 27: ...Start Up 5 26 330985H01 1100 C Box Furnace Chapter 5...
Page 63: ...Maintenance 9 62 330985H01 1100 C Box Furnace Chapter 9...
Page 69: ...Wiring Diagram 10 68 330985H01 1100 C Box Furnace Chapter 10...