background image

Version 1   02/2017

AQ3140 COD Colorimeter PROFORMA User Guide

14

15

Summary of Contents for AQUAfast AQ3140 COD

Page 1: ...AQUAfast AQ3140 COD Colorimeter User Guide Version 1 XCALI 97414 Revision A February 2017 ...

Page 2: ...ries and accumu lators They must not be disposed of in normal domestic waste Because our products include batteries and accumulators in the delivery package our advice is as follows Used batteries and accumulators are not items of domestic waste They must be disposed of in a proper manner Your local authority may have a disposal facility alternatively you can hand them in at any shop selling batte...

Page 3: ... instrument is water proof seal ring E must be in position battery compartment cover B must be fixed with the four screws If the batteries are removed for more than one minute the date and time menu starts automatically when the photo meter is switched on the next time Battery Replacement AAA LR03 AAA LR03 A A A L R 0 3 A A A L R 0 3 D batteries E seal ring C notch F instrument back B battery com ...

Page 4: ... 1 by using deionised water TOC free instead of the sample LR MR 2 ml HR 0 2 ml Replace the cap tightly Invert the vial gently several times to mix the contents Caution The vial will become hot during mixing and digest the vials for 120 minutes in the reactor at a temperature of 150 C Remove the vials from the reactor and allow them to cool down to 60 C or less Mix the contents by inverting each v...

Page 5: ...SET and the instrument returns to the measurement mode On Off Mode Option Mode Mode Zero Test diS Mode SET DATE YYYY 2 sec Mode Zero Test dAtE Option Option CAL METHOD 0 0 0 METHOD METHOD CAL CAL User Calibration Note user calibration Display in calibration mode factory calibration Display in calibration mode After confirming the selection with the MODE key the instrument will show CAL Method Scro...

Page 6: ...ation Mode Technical Data Instrument double wavelength automatic wavelength selection direct reading colorimeter Light source LEDs interference filters IF and photosensor in transparent cell chamber Wavelength specifications of the IF 430 nm 5 nm 610 nm 6 nm Wavelength accuracy 1 nm Photometric accuracy 3 FS T 20 C 25 C Photometric resolution 0 01 A Power supply 4 batteries AAA LR 03 Operating tim...

Page 7: ...contact our Technical Support Specialists Email wai techservbev thermofisher com Within the United States call 1 800 225 1480 Outside the United States call 1 978 232 6000 or fax 1 978 232 6031 For additional product information contact your local authorized dealer Thermo Scientific Orion technical sales represen tative or contact us using the Water and Laboratory Products WLP information on the b...

Page 8: ...Version 1 02 2017 AQ3140 COD Colorimeter PROFORMA User Guide 14 15 ...

Page 9: ...ll Free 1 800 225 1480 Tel 1 978 232 6000 info water thermofisher com Germany Tel 49 6184 90 6000 info water uk thermofisher com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asia thermofisher com Australia Tel 613 9757 4300 in Australia 1300 735 295 InfoWater...

Reviews: