4.
Place the reservoir septa on the CBC, then press firmly to seat the septa.
Note:
The CBC is filled significantly above Fill Line to account for evaporation.
Replace the CBC when the fluid level is at or below the fill line.
Perform these steps if the Cathode Buffer Container has not been installed, if the CBC
on the instrument has expired, or if the buffer level is at or below the fill line.
Install the CBC before you install the cartridge.
See Figure 2 on page 14 for the position of the cathode buffer on the instrument.
1.
Touch , touch Eject plate, then open the instrument door when prompted.
2.
Press the release button on the autosampler to open the lid.
2
1
1
Release button
2
Location of CBC
3.
Insert the cathode buffer in the autosampler with the notch positioned in the
back right.
See Figure 8 on page 20.
Insert the Cathode
Buffer Container
Chapter 11
Maintain the instrument
Install cathode buffer
11
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide
167