The SeqStudio
™
Genetic Analyzer Cathode Buffer Container (CBC) contains running
buffer for capillary electrophoresis. The container has two compartments. The rear
compartment provides the cathode buffer for electrophoresis. The front compartment
is for capillary wash and waste.
The CBC requires a Reservoir Septa.
See “Assemble the SeqStudio
Genetic Analyzer Cathode Buffer Container (CBC)“ on
1
4
5
3
2
Figure 7
SeqStudio
™
Genetic Analyzer Cathode Buffer Container with Reservoir Septa
1
Reservoir Septa
2
Notch (inserted towards the rear right in the autosampler)
3
Fill line (replace the CBC when the buffer is at the fill line)
4
Cathode buffer compartment
5
Waste and wash compartment
1
4
2
3
Figure 8
Autosampler
1
Location of plate or tubes
2
Release button
3
Location of Cathode Buffer Container
4
Location to position the Cathode Buffer Container notch
The cartridge and cathode buffer have radio–frequency identification (RFID) tags. The
instrument reads and tracks:
• Expiry date (shelf life)
• Remaining injections (usage)
• Serial number (cartridge only)
• Lot number (cathode buffer only)
SeqStudio
™
Genetic Analyzer
Cathode Buffer
Container
Radio frequency
identification
Chapter 1
Product information
SeqStudio
™
Genetic Analyzer consumables