If the connection between the instrument and the location that was selected to save the data file
is interrupted, the instrument completes the run. The data file is saved on the instrument. The data
file must be transferred to another location manually (see “Transfer data files at a later time”
).
5.
(Optional)
Select the
Export run results with
checkbox.
For more information about this setting, see “Overview of customized data export”
.
To import a different export setting, see “Import an auto export setting”
.
6.
(Optional)
Edit the run method (see
).
7.
(Optional)
Edit the plate (see
).
8.
Load a plate into the instrument (see
9.
Tap
Start run
, then confirm that a plate was loaded.
The confirmation page can be disabled.
The status dial is displayed. The status dial contains the following information:
• The block temperature
• The elapsed time of the run
• The status of the run
For options during a run, see Chapter 5, “Options for a run”.
Overview of customized data export
The instrument can perform the primary analysis and the secondary analysis of the data, then export
the results.
The instrument has a default export setting.
The export settings can be customized. The instrument will perform the primary and secondary analysis
according to the export setting file.
The export settings are customized in the QuantStudio
™
Design and Analysis Software v2, then
transferred to the instrument. The export settings are in a JSON file format.
The export settings cannot be created or edited on the instrument.
Chapter 3
Create and run plates on the instrument
Overview of customized data export
3
50
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide