Antaris EX-Z2 Specifications
Utilities
Each system requires an AC electrical service, fiber optic cabling for sample
data collection, and clean, dried nitrogen or air for purge and cooling.
Depending on your application, optional Ethernet (for remote computer
and spectrometer control) and analog and digital PLC transmission wiring
also may be used.
Item
Specification
Electrical
AC service
3-wire: live, neutral, earth
ground
Material
8–18 AWG (0.8–8 mm
2
) copper
wire
Tighten input terminals to
2.4 Nm torque
Dedicated
15A
circuit
Input voltage
85-264 VAC
Input current
3.5 A max
Startup current
15 A max
Line frequency
50 Hz to 60 Hz
Line disturbance
Sags, surges or other line
disturbances must not exceed
10% of input voltage (even for a
half cycle)
Noise
< 2 volts (common mode)
< 20 volts (normal mode)
Ethernet
Type
T568A or T568B
Fiber optic
Shielding
Current -carrying conductor
Connections
SMA905
Purge
Gas
Clean, dried air or nitrogen
Filtering
10-micron replaceable filter
Dew point
-70 °C or lower
Pressure and flow rate
Source to purge
4-8 bar gauge
8
scfm
max
Source to cooler
4-7 bar gauge
25
scfm
max
90
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific