Safety Considerations
Fuses
The spectrometer inside the enclosure is protected by two, quick acting
(also known as F-type or fast-blow) fuses. If you need to replace the fuses,
use only fuses that meet the specifications listed below.
Item
Specification
Current 3
A
Voltage 250
VAC
Size
5 x 20 mm
Action
F-type, quick acting
Danger
Avoid shock and fire hazard. Always use exact replacements for fuses.
▲
Danger
Avoid shock hazard. Always disconnect power from the branch circuit that
services the instrument and move spectrometer power switch to the Off
position before checking or changing fuses.
▲
To change a fuse:
1.
Open the enclosure.
Raise keyboard
x4
14
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific