7.3
Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets operating parameters for Relay A; the SP2 option sets
operating parameters for relay B. Since these groups have the same set-up
parameters, they are described together.
7.3.1
Entering Set point 1 (Set point 2) sub-function
SP1
SETUP
HOLD
1)
Enter Advanced Set-up mode
. Push the
ENTER key.
2)
Press
Δ
or
∇
keys
to scroll until upper display
shows SP1 (SP2).
4.00
pH
SP1
SETUP
HOLD
7.3.2
Selecting set point values
This lets you choose the value that will cause your
controller to activate the respective relays.
Lo
SP1
SETUP
HOLD
1)
Follow directions in 7.3.1
to enter Control Relay mode. If you are in this
mode, skip to step 2.
2)
Press ENTER key
. The upper display shows current set point value and
lower display shows SP1 (SP2).
3)
Press
Δ
or
∇
keys
to enter value for Set point 1 (Set point 2).
4)
Press ENTER key
to confirm.
5) Proceed to 7.3.3 step 3, or return to Measurement mode by pressing
Δ
and
∇
keys simultaneously (escape).