Summary of Contents for ALPHA PH 560

Page 1: ...Alpha pH 560 pH Controller pH mV C meter www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...er If you have doubts in the use of the instrument please do not hesitate to contact the nearest Authorised Distributor The information presented in this manual is subject to change without notice as...

Page 3: ...Connecting Temperature Probe 7 Connect Potential Matching Pin PMP 7 Installation 8 Mechanical Dimensions 8 Wall Mount 9 Panel Mount 9 Display Keypad 11 Display Overview 11 Mode Indicators 11 Status A...

Page 4: ...pecifications 36 List of Accessories 38 Troubleshooting 40 General Information 41 Appendices 42 pH Buffer Values at Various Temperatures 42 Graphical Representation of the Function of Hysteresis 43 Ge...

Page 5: ...or pH or ORP and temperature measurement as described in this instruction manual Any other use or use not mentioned here that is incompatible with the technical specifications is deemed inappropriate...

Page 6: ...tions The risk is borne entirely by the user x First disconnect the unit from the power supply and then undo all electrical connections x Remove the unit from the wall panel When the Alpha pH 560 Cont...

Page 7: ...not erased if power supply fails x Menu driven program that simplifies setup x 1 or 2 point calibration with selectable USA or NIST standard pH buffer set x Automatic temperature compensation ATC x M...

Page 8: ...1 Measurement and Control System Remove the screws from the four corners at the back of the pH Controller Remove the back cover The connectors are exposed on the back PCBA as shown in the diagram belo...

Page 9: ...described below Remove screws from the four corners at the back of the pH Controller Remove the back cover Remove the screw located center of the back PCBA Detach the back PCBA from the Controller Tu...

Page 10: ...entific Alpha pH 560 Figure 2 3 Inner Side of Back PCBA Set the J7 jumper to required sensor Pt100 or Pt1000 type Set Jumper J7 www GlobalTestSupply com Find Quality Products Online at sales GlobalTes...

Page 11: ...wire or 3 wire can be connected to the Controller 3 Wire Probe 1 Connect PT100 compensate wire to P Pin 5 of J10 connector 2 Connect PT100 sense wire to Pin 6 of J10 connector 3 Connect PT100 GND wir...

Page 12: ...8 Thermo Scientific Alpha pH 560 Figure 2 5 Frontal View Figure 2 6 Side View Installation Mechanical Dimensions www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 13: ...Thermo Scientific Alpha pH 560 9 Figure 2 7 Figure 2 8 Wall Mount Panel Mount www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 14: ...10 Thermo Scientific Alpha pH 560 Figure 2 9 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 15: ...ode standard pH buffer values are displayed here while measured mV values are displayed in the ORP Calibration mode The two displays indicate function names options settings in Setup mode Refer page 2...

Page 16: ...in its ATC mode Refer Page 26 ERR Appears when an error occurs Electrode annunciator Appears when viewing electrode properties or during calibration error Buffer annunciator Appears in calibration mo...

Page 17: ...etup mode Confirm save setup parameters and numerical values Start Confirm calibration in Calibration mode Select a group of settings in Setup mode Select parameters and increment decrement numerical...

Page 18: ...lower display shows the temperature value The indicator pH or mV at the upper right side of the display indicates the current measurement mode Refer Page Error Reference source not found for switching...

Page 19: ...Thermo Scientific Alpha pH 560 15 Figure 3 13 Menu Overview www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 20: ...ctrode in de ionized water after use and store in electrode storage solution If storage solution is not available use pH 4 01 or 7 00 buffer solution Do not reuse buffer solutions after calibration Co...

Page 21: ...bration Calibration error can be occurred if the measured pH value exceeds this window If you wish to abort the pH calibration Press and keys simultaneously and the pH Controller reverts to pH measure...

Page 22: ...The pH Controller re calculates electrode properties based on the calibration The new slope in mV is shown in the upper display and pH reading at 0mV pH7 00 r offset is shown in the lower display Pres...

Page 23: ...le Press ENT key or CAL key to confirm the reading 4 The calibration is completed The pH Controller re calculates electrode properties based on the calibration The new slope in mV is shown in the uppe...

Page 24: ...se the temperature probe in the buffer solution if ATC mode is enabled Immerse the potential matching pin in the buffer if symmetrical mode is enabled The buffer annunciator appears in LCD Lower displ...

Page 25: ...unciator when the reading is stable Press ENT key to confirm the reading 3 The pH Controller moves to the second calibration point The lower display shows next standard buffer value pH 4 01 U Use and...

Page 26: ...ontroller allows a single point ORP calibration ORP calibration process is available only when the measurement mode is set to ORP Refer Page 32 for switching measurement modes Figure 4 19 ORP Calibrat...

Page 27: ...ller reverts to ORP measurement mode Calibrate temperature probe only if temperature value displayed on the pH Controller is different from that of a calibrated thermometer Refer to page 26 for furthe...

Page 28: ...ccess setup mode LCD shows S SETUP mode indicator and the first page of setup O OFS offset settings P Press or key to access other pages of the setup mode Figure 5 20 Setup Menu NOTE To exit setup mod...

Page 29: ...reading of the pH Controller at the time the sample was taken 2 Measure the pH value of your sample using another calibrated pH meter having its own electrode pH tester hand held meter or bench meter...

Page 30: ...cess liquid is constant and a temperature probe is not available If you disable ATC the pH Controller allows you to set your process temperature P P C and c calibration temperature C C C This allows c...

Page 31: ...are the stabilized temperature reading displayed on the monitor with the thermometer If there is a difference between the two readings offset you can adjust the reading of the monitor P Press or key t...

Page 32: ...ettings define the operating parameters for Relay B Since these two settings have identical setup parameters only SP1 is described here The pH Controller allows you to configure two independent sets o...

Page 33: ...ss ENT key to confirm your hysteresis value N NOTE Hysteresis prevents rapid relay contact switching if the pH reading is fluctuating near the set point Refer to A Appendix IV for a graphical represen...

Page 34: ...the controller type The lower display shows t tyP The upper display shows the last configured controller type OFF or L Ct P Press or key to select the desired controller type O OFF Controller Off Sel...

Page 35: ...key to confirm your selection The pH Controller reverts to C Cntr screen Press or key to access other setup screens or Press and key simultaneously escape to return to measurement mode NOTE The buffer...

Page 36: ...rement operation modes enable disable backlight of the LCD reset the pH Controller to factory defaults Asymmetrical Operation Mode ASY This is the default mode of operation of the pH Controller which...

Page 37: ...LitE The lower display shows last configured backlight selection O On or O OFF Press or key to enable L LitE O On or disable L LitE OFF backlight of the LCD Press ENT key to confirm your selection 3...

Page 38: ...n nO dEF was selected the pH Controller reverts to C Cnfg screen P Press or key to access other setup screens or Press and key simultaneously escape to return to measurement mode If F FCt dEF or C CA...

Page 39: ...Press or key to select viewing electrode properties screen C CdAt Press ENT key to view electrode properties C CdAt 2 Electrode status The electrode annunciator appears in the LCD For pH measurement...

Page 40: ...1000 mV Resolution 1 mV Accuracy 1 mV c Temperature Measuring range 10 0 to 125 0 o C or 14 0 to 257 o F Resolution 0 1 o C 0 1 o F Resolution is 0 1 o F up to 199 9 o F and 1 o F for 200 o F and abo...

Page 41: ...ts display with symbols for status information Back light On Off selectable ELECTRICAL DATA AND CONNECTIONS pH mV Input Screw Terminal 3 5mm pitch Connection terminal 8 pin 2 pin 5 pin terminal blocks...

Page 42: ...ble ECHTPTTSO05B Calibration Solutions Item Description Order Code pH 4 01 buffer solution 480 ml bottle EC BU 4BT pH 7 00 buffer solution 480 ml bottle EC BU 7BT pH 10 01 buffer solution 480 ml bottl...

Page 43: ...ions Item Description Order Code pH 4 01 calibration buffer 500 mL 00654 00 pH 7 01 calibration buffer 500 mL 00654 04 pH 10 01 calibration buffer 500 mL 00654 08 pH 4 01 calibration buffer solution p...

Page 44: ...nsure shield wire is properly connected to pin 7 Slow response a Dirty Oily electrode a Clean electrode Blinking ATC a No temperature probe connection during ATC mode a Ensure temperature sensing cabl...

Page 45: ...truments will not be responsible for any damage resulting from careless or insufficient packing Warning Shipping damage as a result of inadequate packaging is the user s distributor s responsibility P...

Page 46: ...00 6 90 7 04 9 27 10 12 20 4 00 6 88 7 02 9 22 10 06 25 4 01 6 86 7 00 9 18 10 01 30 4 01 6 85 6 99 9 14 9 97 35 4 02 6 84 6 98 9 10 9 93 40 4 03 6 84 6 97 9 07 9 89 45 4 04 6 83 6 97 9 04 9 86 50 4...

Page 47: ...your high set point at pH 10 0 and your hysteresis value at pH 0 5 If your measured value overshoots pH 10 0 the pH Controller s relay activates The actions of the external device will cause the solu...

Page 48: ...Configuration Cntr Controller dEF Default values dEEn De energized En Energized FCt Factory defaults HI High limit HyS Hysteresis LitE Backlight LO Low limit OFS Offset OF d Off delay On d On delay Or...

Page 49: ...nt Thermo Scientific Alpha pH 560 45 Abbreviation Description Set Setting SP1 1st Set point SP2 2nd Set point Sy Symmetrical mode tyP Type Ur Reading is under range www GlobalTestSupply com Find Quali...