S P E C I F I CAT I O N S
Cell Constant
(K)
Conductivity Range
(r1 to r6)
Resolution
0.1, 1.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
1.0, 10.0
Range 1 = 0 – 19.99 uS/cm
Range 2 = 0 – 199.9 uS/cm
Range 3 = 0 – 1999 uS/cm
Range 4 = 0 – 10.00 mS/cm
Range 5 = 0 – 19.99 mS/cm
Range 6 = 0 – 199.9 mS/cm
0.01 uS
0.1uS
1 uS
0.01 mS
0.01 mS
0.1 mS
Temperature Range
0-100
o
C
Temperature Resolution
0.1
o
C
Temperature Relative Accuracy
±
0.5
o
C
Temperature Compensation
Auto (Pt 100) (from 0 to 100
°
C)
Manual (from 0 to 100
°
C)
Temperature Coefficient Factor Setting
0 to 10%
Normalization Temperature Setting
25.0
o
C or 20.0
o
C
Number of calibration points
1
Output
4.0 to 20.0 mA (Galvanically Isolated)
Over- / Under range current output
(Only applicable in the uS range)
‘ON’ Select - 3.8 mA
‘OFF’ Select - 4 mA (under-range) & 20 mA
(over-range)
HOLD current output
‘ON’ Select - 22 mA
‘OFF’ Select - Last measured current value
Display
Custom Dual LCD
Averaging/Stability
(READY indicator)
Yes
Operating Temperature
0 to 100
o
C
Power Requirements
12 to 24 VDC
Load Resistance
Max. 600
Ω
Dimension / Weight
96 x 96 x 66 mm / 210 g