Routine Maintenance
Lamp Maintenance
52
Accela UV/Vis Detector Hardware Manual
Thermo Scientific
Figure 52.
Deuterium lamp assembly alignment pin
4. Securely fasten the assembly in place with the two thumbscrews and standoffs.
5. Connect the lamp lead to the upper terminal in the lamp compartment.
6. Replace the lamp housing and secure it with the screw.
7. Close the front doors of the detector.
8. Plug the power cord back into the power receptacle on the back panel of the detector.
Replacing the Tungsten Lamp
To replace the tungsten lamp, follow these steps:
1.
2.
Installing a New Tungsten Lamp
Removing the Tungsten Lamp
To remove the tungsten lamp
1. Turn the detector off, unplug the power cord from the power receptacle on the back
panel, and allow sufficient time for the lamps to cool before you remove the lamp cover.
2. Open the front doors of the detector.
Alignment pin
CAUTION
Turn the detector off and unplug the power cord from the power
receptacle on the back panel.