Routine Maintenance
Recommended Routine Maintenance
40
Accela UV/Vis Detector Hardware Manual
Thermo Scientific
Recommended Routine Maintenance
lists recommendations for routine maintenance of the UV/Vis detector. Use the table
as a basis for developing your maintenance program in accordance with your company
practices.
Table 5.
Recommended routine maintenance
Procedure
Interval
Cleaning external surfaces
As needed. See
.
Cleaning the flow cell
As needed.
a
.
a
Flushing the flow cell with clean solvent after every use will reduce how often you need to clean it.
Lamp replacement
Deuterium (D2)
Replace the deuterium lamp when the
detector noise reaches an unacceptable
level. See
and
The useful lamp lifetime is
approximately 2000 hours at a diode
array scan rate of 20 Hz.
Tungsten (W)
Every 2500 hours or as required. See
.