6
Routine Maintenance
Replacing the Fuses
Thermo Scientific
Accela PDA Detector Hardware Manual
95
Replacing the Fuses
You can operate the PDA detector at 100/115V or 230V. However, you must make sure that
the appropriate fuses are installed (see
).
Y
To replace the fuses
1. Turn off the power to the detector and unplug the detector from line power.
2. Insert the tip of a narrow-blade screwdriver into one of the two openings on the exterior
surface of the power entry module, and then apply leverage to gain access to the power
entry module’s door (see
Figure 78.
Loosening the power entry module’s door
Table 4.
Fuses
Line power voltage
Fuse size
100/115V
T3.15 A
230V
T1.6 A
CAUTION
To avoid an electrical shock, before you replace the fuses, turn off the
power to the detector and unplug the detector from line power.