2
Accela 600 Pump or Accela 1250 Pump Installation
Solvent Line Connections
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
33
shows the Hot Surface caution symbols on the column oven surface.
Figure 19.
Hot surface cautions on the surface of the Accela Autosampler column oven
To connect the high-pressure solvent line to the heat exchanger inlet
1. Ensure that the column oven compartment is at room temperature and that the pump
flow is off.
2. Route the stainless steel tubing through the access slot in the autosampler drip tray.
3. Connect the tubing to the heat exchanger inlet as follows:
a. Using a two-piece, stainless steel fitting, connect the tubing to the free end of the
union connected to the heat exchanger inlet.
b. Using a 1/4 in. open-ended wrench, tighten the nut.
shows the high-pressure connection between the Accela Autosampler and the
Accela 600 or 1250 Pump.
CAUTION
To prevent personal injury, before installing the LC column or connecting the
pump to the autosampler, ensure that the column oven compartment is at room
temperature. The column oven compartment can reach temperatures as high as 95
C
(203
F).
shows the connections for the latest version of the Accela Autosampler
heat exchanger.
Hot surface caution
symbols