Troubleshooting
Diagnostic Tests
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
111
9. If your flow rate is unstable or inaccurate, see
.
Testing the Residual Pulsatio
n
This test measures the pump pulsation.
To measure the pump pulsation
1. Connect a 40 bar (~580 psi) restrictor cartridge to the LDA outlet. Align the restrictor so
that the flow is in the direction of the arrow marked on the restrictor (see
2. Fill a mobile phase reservoir with HPLC-grade methanol and connect it to inlet
channel A.
3. Open the purge valve.
4. Set the pressure unit to bar.
5. Verify that the pressure is 0.
6. Set channel A to 100%.
7. Set the flow rate to 500 μL/min.
8. Start the pump, wait 1 minute, and then close the purge valve.
9. Wait 5 minutes to allow the pump pressure to equilibrate.
10. Observe the pressure graph for pressure fluctuations. The pressure standard deviation
should be less than 1 bar (14.5 psi).
Table 13.
Reasons for unstable flow rates
Possible reason for unstable flow rate
Correction
Damaged or dirty check valve
Clean the check valves. See
.
Worn out primary piston seals
Replace the seals. See
.
Unstable load (column)
Use a backpressure restrictor or another
column.
Tip
For information on viewing the pump’s status and starting and stopping the
solvent flow from the pump, refer to the data system Help.