Speed/Tilt

Time

Mode

Speed

Tilt

Run

Run

Stop

rpm/degree

4.4 Loads, Angle and Speed

Asymmetric loading or unloading may affect the 
normal operation of the instrument and may 
even cause unpredictable damage.

Section 5  Operation

This chapter covers the control panel and its operation.

5.1 Control Panel

.  Speed/Tilt setting knob: Turn the  

      knob clockwise or counterclockwise 
      are used to increase/decrease the set 
      speed/Tilt of instrument.

.  Speed/Tilt display window: The 

      window displays set speed/tilt 
      (when the instrument is in standby)  
      or current speed/tilt (when the 
      instrument is running).

.  Time display window: The 

      window displays cumulative time (in 
      continuous mode) or remaining time 
      (in timer mode). The range of time 
      displayed is 0s to 9999 minutes. 

.  Time setting knob: Turn the  

      knob clockwise or counterclockwise 
      are used to increase/decrease the set 
      time of instrument.

.  Run/Stop button: Start or stop the 

      instrument.

.  Run indicator: The light is on when  

      the instrument is running and off   
      when the instrument is in standby.

.  Mode button: Switch modes between 

 

     Speed and Tilt settings.

.  Tilt indicator: The light is on when 

 

     the instrument is in Tilt Adjustment 

 

     Mode.

. Speed indicator: The light is on when 

 

     the instrument is in Speed 

       Adjustment Mode.

. Gear I:When the switch is in “gear 1", 

      the instrument power is on.
    . Gear O:When the switch is in “gear O”,  
      the instrument power is off. 

7      Section 4  |  Overview

Digital Platform Rocker      8

Warning:

 It may affect the maximum 

speed of the instrument if placing the 
cell culture flask in the corner or on 
one side of the tray. If asymmetrical 

placement is required by the experiment, 
do not set to max speed instead  change
the speed gradually.

24°

Maximum Speed 

(rpm)

Placement

Tilt Angle

Maximum Weight

(Kg)

Centered

Centered

70

70

2.0

2.0

24°

Centered

Centered

50

50

4.0

4.0

1°-24°

1°-24°

Maximum Speed 

(rpm)

Placement

Tilt Angle

Part of Accessory

Maximum Weight

(Kg)

Centered

Centered

70

70

1.0

2.0

Double Platform

Double Layer Platfrom

Single Platform

11

11

Power Switch

Summary of Contents for 88882013

Page 1: ...fisher com Find out more at thermofisher com In Canada For customer service call 1 800 234 7437 To fax an order use 1 800 463 2996 To order online thermofisher ca Operating Manual Revision A 04 01 2021 2021 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified ...

Page 2: ...rview Specifications Environmental Conditions Safety Instructions Loads Angle and Speed Section 5 Operation Control Panel Installation of Two Trays Installation Settings Section 6 Safety Tips and Maintenance Section 7 Troubleshooting Section 8 Optional Accessories Section 9 Warranty Information Section 10 Compliance 16 Contents 1 2 2 2 3 3 4 4 5 5 5 6 7 8 8 9 10 10 12 13 14 15 ...

Page 3: ... power switch and for allowing the instrument to run without impediment Do not install the stirrer near machinery generating high frequency noise Avoid installation close to high frequency welding machine sewing machine or mass SCR controller Do not inject any liquid and inflammable things inside of product Do not pour water or put liquid on the top of the product when cleaning Disconnect the main...

Page 4: ...ed during transit After unpacking check that all unit parts and accessories are as listed below Contact us or the agent from which you purchased the unit if any components were omitted 3 Section 3 Inspection and Installation Digital Platform Rocker 4 General Power Adaptor US Plug CN Plug EU Plug UK Plug Catalog Number 88882013 1 10 N A N A N A 6 1 1 88882014 10 1 N A 6 1 1 1 1 Figure Dimpled Rubbe...

Page 5: ...orm Rocker in a hazardous atmosphere or with hazardous materials for which the unit was not designed Also the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by the manufacturer or are used in a manner not specified by the manufacturer CAUTION To avoid electrical shock completely cut off power to the unit by disc...

Page 6: ...r stop the instrument Run indicator The light is on when the instrument is running and off when the instrument is in standby Mode button Switch modes between Speed and Tilt settings Tilt indicator The light is on when the instrument is in Tilt Adjustment Mode Speed indicator The light is on when the instrument is in Speed Adjustment Mode Gear I When the switch is in gear 1 the instrument power is ...

Page 7: ... Press the Time setting knob To decrease time rotate the knob counter clockwise to increase time rotate the knob clockwise then press the knob with in 5sec to confirm the setting the digital tube will stop flashing and the time setting is complete the instrument is in timer mode Speed Settings 1 Press the button and the Speed indicator light will be on 2 Press the Speed Tilt setting knob To decrea...

Page 8: ...arning The operator is not permitted to gain access using a tool to a part which in normal use may be hazardous live Digital Platform Rocker 12 Run and Stop Press button and the instrument will start running with the specified settings and the Run indicator light will be on The Time dispaly window will show the cumulative time Continuous Mode or remaining time Timer Mode and the Speed Tilt display...

Page 9: ...rews Place instrument at a flat and firm suface Fasten housing screws Other Keep record for maintenance Shut down and restart useless please contact Thermo Scientific Err1 Operational obstruction Section 8 Optional Accessories Description Cat No Max Qty Figure Rubber Mat DoubleLayerTrayw 6screws 88882102 88882103 88882101 88882148 1 1 6 Rubber Strips 6 Ea pack Dimpled Rubber Mat General Power Adap...

Page 10: ...ice and special application Outside the USA please contact local Thermo Technical Services Department or local distributor for warranty information Section 10 Compliance 1 European Union The European voltage models of this product meet all the applicable requirements of the European Directives and therefore display the CE Marking These Directives are captured in the EU Declaration of Conformity wh...

Page 11: ... the Control of Pollution Caused by Electronic Information Products ACPEIP A label of conformance such as one of the following may be found on the product 5 End of Life Care Some considerations and suggestions are listed below for proper disposal of this product While addressing these actions for safe recycling and disposal please follow all guidelines Safety Data Sheets SDS or regulations applica...

Reviews: