restoration. The display windows will 
flash. Press any button to stop flashing.

4.4 Accessory Installation 

1. 

(figure 1),  and 

r

2. Split the Fixed Tube Rack and the 

holder (88882117, 88882118, 
88882119), put the holder on tray and 
fix it with the 4 sunk screws 

.

3. Vertically clip the Fixed Tube Rack into 
    the holder and make sure there is no
    gap between the rack and the holder  

.

4. Insert the test tube.

Note:

 When using single test tube, it is 

recommended not to set the speed too 
high because the imbalanced load may 
damage the instrument.

Loosen the screws of the tray with a   

    cross screwdriver 
     eplace the standard tray with platform 
    large (figure 2).

    
    
    

(figure 3)

    (figure 4)

9      

 |  

Section 4  Operation

Finish Operation

 

After the operation is finished, press the 
power switch at the back right side of 
the instrument and put it into the “

O

” 

state. Unplug the instrument and store 
the instrument according to the storage 
guide.

Alarm System

Exceed the speed limit: After entering the 
operation mode for 5s, if the actual 
speed of the instrument is 0 or more 
than 1200rpm, the instrument will alarm 
immediately.

Err1: After entering the operation mode 
for 10s, and the instrument runs steadily 
(running stability means that the actual 
speed of the instrument is within the set 
value of ±10rpm). If the speed exceeds 
the set speed of ±10rpm, the instrument 
stops running and buzzes to alarm, Run 
indicator light goes off,  and "Err1" is 
displayed in the Speed display window.

End of timer: The instrument buzzing 
alarm, the "End" is displayed in the Speed 
display window.

When the instrument alarms, press any 
key, the instrument is back to the 
standby mode.

Power Recovery

If the power supply is cut off suddenly 
while the instrument is in operation, the 
unit will automatically run at the 
previously set parameter upon power

figure 1

figure 2

figure 3

figure 4

Digital Microplate Shaker      10

4.2 Installation

1. Connect all the components according 
    to the figures shown  on page 4 of this 
    manual. Use grounded power outlet.
2. Press the power switch "I" side and 
    switch on the instrument.

4.3 Settings

Time Settings

1. Continuous mode
Press the “      ” or “      ” arrow button 
below the Time display window. When 
the number shown on the Time display 
window starts flashing, press “      ” 
arrow button to decrease the time to 
00:00 and then release the button. The 
time setting is finished after the number 
shown on the Time display window 
flashed twice.

2. Timer mode
Press the “      ” or “      ” arrow button 
below the Time display window. When 
the number shown on the Time display 
window starts flashing, press “      ” or 
“      ” arrow button to increase or 
decrease the time value. Release the 
button when the time shown on the 
Time display window reaches the set 
value. The time setting is finished after 
the number shown on the Time display 
window flashed twice.

Speed Settings

Press the “      ” or “      ” arrow button 

Run

Stop

Run

Stop

below the Speed display window. When 
the number shown on the Speed display 
window starts flashing, press “      ” or      
“      ” arrow button to increase or 
decrease the speed value. Release the 
button when the speed shown on the 
Speed display window reaches the set 
value. The speed setting is finished after 
the number shown on the Speed display 
window flashed twice. 

Note:

 

Press the “      ” or “      ” arrow 

button for a longer time to accelerate the 
setting.

Run and Stop

Press “      ” button and the instrument 
will start running with the specified 
settings and the Run indicator light will 
be on. 
The Time display window will show the 
cumulative time (Continuous Mode), or 
remaining time (Timer Mode) and the 
Speed display window will show the 
current speed.
Press “      ” button again and the 
instrument will slow down until it stops. 
The instrument will then be in standby 
and the two display windows will show 
the set values.

Note: 

To ensure shaking operation 

smooth and steady, it may take 1minute 
for the microprocessor control system to
accelerate the tray to the set speed. 

Summary of Contents for 88882005

Page 1: ...hermofisher com In Canada For customer service call 1 800 234 7437 To fax an order use 1 800 463 2996 To order online thermofisher ca For Research Use Only Not for use in diagnostic procedures 2019 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified Operating Manual Revision A 09 03 2019 ...

Page 2: ...Overview Specifications Environmental Conditions Safety Instructions Capacity and Speed Section 4 Operation Control Panel Installation Settings Accessory Installation Section 5 Safety Tips and Maintenance Section 6 Troubleshooting Section 7 Optional Accessories Section 8 Warranty Information Contents 1 2 3 3 4 4 5 5 5 6 7 8 8 9 9 10 11 12 13 15 ...

Page 3: ...g can happen Do not assemble repair modify on your own The product may not work well and electric shock is possible with changes in the efficiency of the product Also this will void the warranty Indicates a hazardous situation which if not avoided may result in minor or moderate injury Do not put heavy things on the power cord Do not put the machine on the cord It may take off the wire coating and...

Page 4: ...it Then unpack the unit Check carefully to see if there were damages incurred during transit After unpacking check that all unit parts and accessories are as listed below Contact us or the agent from which you purchased the unit if any components were omitted 3 Inspection Section 2 and Installation Screw Driver 1 1 12 12 Tray 96 Microplate Clamp 6 Installed well M3 6 Screw Installed Main Body Powe...

Page 5: ... operating the Digital Microplate Shaker Digital Mircopalte Shaker DO NOT immerse the unit for cleaning DO NOT operate the unit if it shows WARNING DO NOT use the in a hazardous atmosphere or with hazardous materials for which the unit was not designed Also the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by t...

Page 6: ... Speed display window The window displays set speed when the instrument is in standby or current speed when the instrument is running Time display window The window displays cumulative time in continuous mode or remaining time in timer mode Time setting buttons UP DOWN arrow buttons are used to increase decrease the set time of the instrument Speed setting buttons UP DOWN arrow buttons are used to...

Page 7: ...ital Microplate Shaker 10 4 2 Installation 1 Connect all the components according to the figures shown on page 4 of this manual Use grounded power outlet 2 Press the power switch I side and switch on the instrument 4 3 Settings Time Settings 1 Continuous mode Press the or arrow button below the Time display window When the number shown on the Time display window starts flashing press arrow button ...

Page 8: ...urface of the instrument Place the instrument in a ventilated and dry environment for 24 hours before reuse If the instrument is not functioning after drying for 24 hours please contact the manufacturer Warning Disassembling Assembling without a qualified professional s guidance may cause malfunctioning of the instrument Section 5 Safety Tips and Maintenance Section 6 Troubleshooting Error Cannot ...

Page 9: ...er Adaptor w AUS CN plug General Power Adaptor w EU plug General Power Adaptor w UK plug Platform Large Assemble For 6 x Microplates 96 well Microplate clamp Digital Microplate Shaker 14 88882130 88882131 88882132 Screw GB T819 1 M4 8 Platform installation Screw GB T819 1 M5 10 Tube rack installation Screw GB T9074 4 M3 6 96 well Microplate clamp installation Dimensions 292x275x14mm 292x275x25mm 1...

Page 10: ...the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB destination THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLI...

Reviews: