FH15 & FH30 Compact Pump System Operating Manual 2-1
Thermo Scientific
Section 2
Installation and Setup
Use only two-stop tubing sets to ensure optimum performance. Use
of other tubing sets may void applicable warranties.
Appendix A provides a list of tubing sets which will work efficiently with
the FH15 & FH30 Compact Pump System. This list includes tubing
diameters, flow rates in mL/min and the maximum flow rates per
revolution at 20 rpm and 100 rpm. Tubing is listed by part number. For
best results, when using two tubes (two-channel models only), select tubes
that provide flow rates within a 10:1 range. (Tube diameter ratios should
be within approximately 3.5:1.)
WARNING: Before installing tubing into the Pump Head
remove power or shut off the drive. Fingers or loose clothing
could be caught in the rollers.
For dual or two-channel Pump Systems, Silicone tubing should not be
used with TYGON® or PHARMED® BPT.
1. Place the Power On/Direction Switch, Figure 1-2, in the center
(Off ) position.
2. Push the Occlusion Bed Latch, Figure 2-1, to the right to release the
occlusion bed from the rotor.
Figure 2-1.
Occlusion Bed in Unlatched Tube Loading Position
Selecting Tubing Size
Installing Tubing in
Pump Head
Setup
Occlusion Bed Latch
Occlusion Bed in
Unlatched (Tube
Loading) Position
Rear Tube
Retainer
Slots
Front Tube
Retainer
Slots