Servicing
Pump Replacement
7-10
Model 49
i
Instruction Manual
Thermo Fisher Scientific
3. If either fuse is blown, replace both fuses.
4. Insert fuse drawer and reconnect power cord.
Pump Replacement
Use the following procedure to replace the pump (see
). To
rebuild the pump, see “Pump Rebuilding” in the “Preventive Maintenance”
chapter.
Equipment Required:
110V pump
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect pump power line from AC PUMP connector on the
measurement interface board.
Figure 7-4.
Replacing the Pump
3. Remove both lines from pump.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Pump
Pump Bracket
Captive Screws (4)
Shock Mounts (4)
Floor Plate
Pump Mounting Screws (2)
(Bottom)