Optional Equipment
Connecting External Devices
9-4
43iQ Instruction Manual
Thermo Scientific
The Analog I/O Board consists of:
●
4 Isolated Analog Voltage Inputs, Input Voltage Range: 0–10 V
●
6 Isolated Analog Voltage Outputs, Three Ranges: 0–1.0 V, 0–5.0 V,
0–10 V
●
6 Isolated Analog Current Outputs, Two Ranges: 0mA–20mA, 4mA–
20mA
lists the analog voltage inputs are used to monitor four external
0–10V signals.
Figure 9–4.
Analog Voltage Inputs
Table 9–3.
Analog Voltage Inputs Assignment
Terminal Number
Signal Name
1
Analog In 1
2
Analog GND
3
Analog In 2
4
Analog GND
5
Analog In 3
6
Analog GND
7
Analog In 4
8
Analog GND
Analog I/O Board
Analog Voltage Inputs