Servicing
Permeation Oven Replacement
Thermo Scientific
43iQ Instruction Manual
7-45
4.
Push in fitting head evenly with fingers towards fitting body.
Figure 7–51.
Finger Push and Release Tubing
5.
Pull tube outwards to release. If using the tool provided:
a.
Adjust tool size to 1/4 tube as indicated on the back side.
b.
Tool edge should be pushed straightforward into the edge of the
release button along with the tube in the axial fitting direction.
c.
After inserting, grasp handle tightly and insert the end of the tubing
to the stroke end.
Note
Insert firmly to the guard against an accidental tube release.
▲
d.
After inserting end of tube, relax your grip on the tool. Returning
force of spring releases the tube.