Servicing
DMC Optical Bench Removal
7-34
43iQ Instruction Manual
Thermo Scientific
5.
Using #1 Phillips drive, unfasten three #4-40 pan head screws.
Figure 7–38.
Removing the PMT Cover
6.
Slide cover and grommet back.
Note
Wear lint-free gloves to unplug PMT from base.
▲
7.
Using #2 Phillips drive, unfasten two #6-32 pan head screws.
8.
Replace PMT as needed and assemble in reverse order.
Figure 7–39.
Replacing the PMT