Servicing
DMC Optical Bench Removal
Thermo Scientific
43iQ Instruction Manual
7-33
Use the following to replace the photomultiplier tube (PMT).
Equipment required:
Hex drive, 5/32
Phillips drive, #1 and #2
Lint-free gloves
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Unplug all electrical as shown (
).
3.
Using 5/32 hex drive, unfasten two #10-32 socket cap head screws.
4.
Pull photomultiplier housing upwards.
Figure 7–37.
Removing the PMT Assembly
Photomultiplier
Tube (PMT)
Replacement