Servicing
DMC Optical Bench Removal
Thermo Scientific
43iQ Instruction Manual
7-29
Use the following to remove the DMC optical bench from the instrument.
●
Optical Bench Assembly Removal
●
PMT Replacement
●
Flasher Pack Replacement
●
Flasher Pack Lamp Replacement
●
Photo Lamp Detector Board Replacement
●
Kicker Assembly Replacement
Use the following procedure to remove the optical bench assembly.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Unplug two DMC cables.
3.
Disconnect all plumbing.
4.
Unfasten six captive hardware.
DMC Optical
Bench Removal
Optical Bench
Assembly Removal