Servicing
DMC Pressure and Flow Board
Thermo Scientific
43iQ Instruction Manual
7-17
Use the following to replace the DMC pressure and flow board.
Equipment required:
Hex drive, 7/16
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Unplug cables from the pressure and flow board.
Figure 7–20.
Flow Pressure Board, Disconnect DMC Cable
3.
Disconnect plumbing.
4.
Using 7/16 hex drive, unfasten four #6-32 socket cap head screws.
DMC Pressure
and Flow Board