Temperature Ranges

Refrigerator: 3° to 13°C (37° to 55°F)

Freezer: -20° to -10°C (-4° to 14°F)

Temperature Stability

@ 4 °C ±4 °C

@ -20 °C ±4 °C

Electrical Requirements

See dataplate attached to unit.

5

Performance Characteristics

Summary of Contents for 3552-10A

Page 1: ...Installation and Operation Manual Thermo Scientific General Purpose and Flammable Storage Laboratory Refrigerators and Freezers ...

Page 2: ... 2010 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 3: ...lammable Materials Storage Refrigerators and Freezers Cool Lab General Purpose 3767A 3767 1A 3766A 3766 1A Flammable Material Storage 3552 10A 3552 11A 3552 11ACN 3566 10A 3566 11A 3566 11ACN 3560 10A 3560 11A 107774 Rev A ...

Page 4: ...cs 5 Temperature Ranges 5 Temperature Stability 5 Electrical Requirements 5 Installation 6 Selecting a Location 6 Leveling the Unit 6 Shelves 6 Electrical Connection 6 Operation 8 Initial Startup 8 Temperature Control 8 Condensate Disposal 8 Material Compatibility 8 Troubleshooting 10 Maintenance 11 Cabinet Cleaning 11 Condenser 11 Warranty 12 ...

Page 5: ...y This product is not intended nor can it be used as a ster ile or patient connected device In addition this appara tus is not designed for use in Class I II or III locations as defined by the National Electrical Code unless otherwise noted Safety Information Warning Warnings alert you to a possibility of personal injury Caution Cautions alert you to a possibility of damage to the equipment Note N...

Page 6: ...ribe such external evidence of loss or damage may result in the carrierʼs refusing to honor your damage claim The form required to file such a claim will be supplied by the carrier Concealed Loss or Damage Concealed loss or damage refers to loss or damage which does not become apparent until the merchandise has been unpacked and inspected Should either occur make a written request for the carrierʼ...

Page 7: ...mperature Ranges Refrigerator 3 to 13 C 37 to 55 F Freezer 20 to 10 C 4 to 14 F Temperature Stability 4 C 4 C 20 C 4 C Electrical Requirements See dataplate attached to unit 5 Performance Characteristics ...

Page 8: ...ront to back by adjusting the supplied leveling feet The leveling feet can be adjusted so the front of the unite is slightly higher than the rear this will allow the door to close easily if it is left halfway open Shelves Shipped inside each cabinet are four fixed shelves which also serve as part of the evaporator coil on freezer mod els Electrical Connection Determine the total amount of current ...

Page 9: ...t checked by a quali fied electrician to verify that they are appropriate for the power requirements of the product that they match the supplied plug and that they are properly grounded and have over current protection WARNING DO NOT under any circumstances cut or remove the third ground prong from the power cord DO NOT use a two prong adapter plug Where a two prong wall receptacle is encountered ...

Page 10: ...ty The interior cabinet of this unit is constructed of epoxy coated steel Care must be exercised when determining which chemicals may be stored in the refrigerator and freezer sections and which type of storage materials should be employed ABS Plastic deteriorates when exposed to but not limited to the following Aliphatic Hydrocarbons Aromatic Hydrocarbons Fully and Partially Halogenated Hydrocarb...

Page 11: ...s with mixtures of organic compounds cannot always be predicted from their compatibility with each of the compounds For example some aluminum alloys are corroded severely in mixtures of carbon tetra chloride and methyl alcohol even though they are resist ant to each compound alone This information is taken from Corrosion and Corrosion Protection Handbook Second Edition Published by Marcel Dekker I...

Page 12: ...circuit Check fuse or circuit breaker at breaker breaker box Runs Continuously Frost buildup on refrigeration coils Defrost unit Clicking Sound The compressor is equipped Unplug unit for 1 hr Plug in and try with a thermal protector running again If the unit doesnʼt This device shuts off the begin to run call for service compressor when it becomes too hot A clicking sound occurring about every 30 ...

Page 13: ...er is recommended for cleaning the interior and exterior of the cabinet All surfaces should be rinsed and thoroughly dried Condenser The condenser coil is located behind the sides and rear exterior panel These surfaces may be warm to the touch This is necessary to the operation of the refrigeration equipment and is normal 11 Maintenance Caution Do not use any type of abrasive such as steel wool or...

Page 14: ...authorized service representa tive ii misuse including use inconsistent with written operating instructions for the product mishandling contamination overheating modification or alteration of the product by any customer or third party or iii use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific prod ucts Heating elements because of their susce...

Page 15: ... finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fisher Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o rici...

Page 16: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Page 17: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 107774 Rev A ...

Reviews: