Optional Equipment
Permeation Oven
Thermo Fisher Scientific
Model 146
i
Instruction Manual
9-7
3. From the Permeation Oven menu, in the Service menu, select Cal Gas
Thermistor >
Water Bath
. Enter the thermistor temperature from the
thermometer and press
).
Figure 9-2.
Cal Gas Therm Bath Screen
4. Remove the thermistor from the water bath, dry, and replace into the
permeation tube oven.
5. Make sure the charcoal scrubber is connected to the ZERO bulkhead on
the rear panel.
6. Wait for the permeation gas temperature reading to stabilize.
Setting Temperature with Known
Resistance
Use the following procedure to set the gas temperature with an accurate
known resistance.
1. Remove the thermistor from POJ1 on the measurement interface board.
1. Connect a 4.369 K
Ω
resistor across pins 1 and 2 of POJ1 (use a
resistance substitution box and an accurate meter, if necessary).
2. From the Permeation Oven menu, select
Cal Gas Thermistor
, then
select known resistor (
).
SAMPLE 12:34 ALARM
x
CAL GAS THERM (BATH):
CURRENTLY: 45.80
o
C
SET TO: 45.0
0
o
C
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...