Preventive Maintenance
Ozonator Capillary Replacement (Ozonator option)
5-6
Model 146
i
Instruction Manual
Thermo Fisher Scientific
•
Check the ends of the capillary for an obvious obstruction and
remove the obstruction.
•
Clear any internal blockage using an appropriate wire for the
particular capillary. Refer to
.
•
If a restriction cannot be cleared, replace the capillary. Refer to the
appropriate capillary replacement procedure that follows.
Refer to
to locate these capillaries within
the instrument. If you need the capillary part number, refer to the
Ozonator Capillary
Replacement
(Ozonator option)
Use the following procedure to replace the ozonator capillary
).
Equipment Required:
Ozonator capillary
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unscrew the knurled fitting, separate the components, and remove the
capillary (
Table 5-1.
Capillary Cleaning Wire Sizes
Capillary
Wire Diameter
Ozonator
8 mil
Photometer
8 mil
Pressure Gauge
8 mil
Pump
20 mil
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...