Chapter 2 Installation
2-5
Figure 2-3.
Scrubber Base and Top Plates
Be sure to properly seat all of the canisters, even if one or more do not contain packing.
Failure to properly install the canisters will introduce leaks or ambient air into the system.
Note
. If Drierite is used as a desiccant, it should be in the front canister so that it can be
monitored through the view port in the front panel. When this desiccant turns pink, it
should be changed.
O-Rings