Unpacking and Installation
Thermo Scientific Laboratory Refrigerators and Freezers
Installation and Operation Manual
15
5 Unpacking and
Installation
5.1 Shipping Carton
This should be inspected upon delivery. When received, carefully examine
for any shipping damage before unpacking. If damage is discovered, the
delivering carrier should both specify and sign for the damage on your copy
of the delivery receipt.
Open the carton carefully making certain that all parts are accounted for
before packaging materials are discarded. After unpacking, if damage is
found, promptly report it to the carrier and request a damage inspection
promptly.
IMPORTANT
Failure to request an inspection of damage within a few
days after receipt of shipment absolves the carrier from any liability for
damage. You must call for a damage inspection promptly.
5.2 Unpacking
Use the list below when unpacking to verify that the complete unit has been
received. Do not discard packing materials until all is accounted for.
The following items are included in the shipment:
5.3 Location
Locate your unit in the most convenient place and near a grounded electric
outlet. If units are placed on a counter-top, the front should be 3" or more
back from the edge to avoid accidental tipping of the unit.
If possible, locate your unit out of direct sunlight and away from heat
sources such as a radiator, stove or heat duct.
5.4 Clearance
Position the unit so that there is at least clearance as per the table below. It is
recommended that even more space be left at the back of the unit in case
maintenance is required.
Refrigerator/Freezer
Operation Manual
057-196-00
Volt Warning Tag
528-609-00
Inspection Tag
528-028-00
Model
Top, in.
Rear, in.
Sides, in.
05LREETSA/05LREETSV/10LCEETSV/10LCEETSA/05LCEETSA/05LCEETSV/
04LFEETSA
2
2
2
05EREETSA/10ECEETSA/10ECEETSV/10FCEETSA/10FCEETSV/05FREETSA
3
12
3
12LCEETSA
4
4
4
02LFEETSA
5
5
5