5

3.5

SECURING THE SAMPLES

When all samples are loaded, rotate the spoke plate by turning the rotation knob coun-
terclockwise so that the retention spokes are centered above each sample tube cap.
Hand-tighten the three securing knobs while pushing down on the rotation knob.

WARNING:

Samples must be secured properly before running the FastPrep

®

Instrument. Overtightening of FastPrep

®

knobs may result in damage to FastPrep

®

spokeplate and FastPrep

®

knobs.

Close and lock cover of FastPrep

®

FP120.

3.6

RUNNING THE UNIT

Start run by setting switch (RUN/OFF/CHANGE) to RUN.

If display flashes “CC”, the cover is not properly secured. Firmly press down cover to lock.
(As a safety feature, the FastPrep

®

Instrument will not operate when cover is not properly

locked.)

To display speed during the run, set switch (SPEED/TIME/COVER) to SPEED.
To display time remaining in the run, set switch (SPEED/TIME/COVER) to TIME.

Note:  When SPEED is selected on switch A, “AA” is displayed during motor acceleration.
When the motor reaches its preset speed, the set speed value is displayed.  If the set speed
is not reached within 10 seconds of starting a run, an error message appears (E.3.), and
the run is automatically aborted.  During motor deceleration at the end of a run,  “dd” is dis-
played.  If the motor does not decelerate to a complete stop within 10 seconds, error mes-
sage E.4. appears.  Call Qbiogene immediately if an error code appears.

At the end of a run, the motor brakes to a stop automatically, and the display reads “EE”
to indicate that the run has ended.  Set switch (RUN/OFF/CHANGE) to OFF to unlock the
cover (the “EE” display stays on for 2 seconds, then returns to the preset value).  Open
cover and remove sample tubes.

Warning:  Use caution when removing samples; sample holder may become hot after
the run.

Summary of Contents for FastPrep FP110

Page 1: ...Inc North America 2251 Rutherford Road Carlsbad CA 92008 Phone 800 424 6101 Fax 760 918 9313 Qbiogene Inc Europe Parc d Innovation BP72 67042 Illkirch Cedex France Phone 33 0 388 675 425 Fax 33 0 388...

Page 2: ...tting Target Time 4 3 3 Example of Manual Run 5 3 4 Example of Auto Run 6 3 5 Drying Rate 6 3 6 Other Topics 7 Section Section Title Specifications Page 4 0 Specifications 8 Section Section Title Acce...

Page 3: ...675 425 Europe or visit the FastPrep Users Resource Group at http www qbiogene com 2 0 INSTALLATION Unpacking Carefully remove the FastPrep Instrument and accessories from the shipping carton Compare...

Page 4: ...ble power source and the RUN OFF CHANGE switch on the front control panel is turned OFF see Figure 1 Below Once the main power is ON press COVER SPEED TIME COVER switch to release the lid 2 0 INSTALLA...

Page 5: ...it returns to the OFF position Note While setting run speed the display counts forward in increments of 0 5 m sec within the operating range of 4 0 6 5 m sec B Select desired time of run The units di...

Page 6: ...ke sure that their screw caps are tight Load sample tubes into the sample holder Whenever possible position the tubes symmetrically The FastPrep Instrument is designed to be used exclusively with the...

Page 7: ...switch A SPEED TIME COVER to SPEED To display time remaining in the run set switch A SPEED TIME COVER to TIME Note When SPEED is selected on switch A AA is displayed during motor acceleration When th...

Page 8: ...peed Construction Anodized aluminum sample holder Polypropylene sample retention spoke plate Chemical resistant painted surfaces Dimensions 11 in x 15 75 in x 12 in W x D x H 28 0 cm x 40 0 cm x 30 5...

Page 9: ...of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous materials you must pack and l...

Page 10: ...PENDIX 1 MAINTENANCE CLEANING The cell disruption process during a FastPrep Instrument run is caused by the collision of matrix and sample within the FastPrep Instrument sample tube The rate of collis...

Page 11: ...interlock override handle is visible underneath situated toward the front The override mechanism is a flat metal handle with release cord wound around attached to the undersurface by two Phillips hea...

Page 12: ...104 3003 00 Rev B 2000 Thermo Savant...

Reviews: