System Shutdown, Startup, and Reset
Shutting Down the System in an Emergency _______________________________________
3-2
____________________ LCQ Deca XP Plus Hardware Manual ________________
3.1 Shutting Down the System in an
Emergency
If you need to turn off the MS detector in an emergency, place the main power
circuit breaker switch (located on the power panel on the right side panel of
the MS detector) in the Off (O) position. See Figure 3-1. This turns off all
power to the MS detector, including the vacuum pumps. Although removing
power abruptly will not harm any component within the system, this is not the
recommended shutdown procedure to follow. Refer to the topic Shutting
Down the System Completely, for the recommended procedure.
To turn off the LC, autosampler, and computer in an emergency, use the
On/Off switches on the LC, autosampler, and computer, respectively.
Figure 3-1.
Power panel, showing the Reset button, electronics service switch, and the
main power circuit breaker switch
Main Power
Circuit Breaker Switch
Electronics
Service Switch
Reset Button