Colorimeters come pre-calibrated, however a user can
check and adjust calibration if desired.
Always use standard in recommended range:
1. Hold the
mode
key while turning on the
colorimeter with the
power
key.
2. Fill clean dry cuvette to mark with standard
and cap.
3. Place cuvette in colorimeter and press
zero•test
key.
4. Remove cuvette, remove cap, and add tablet or
reagent. Crush and stir with stir rod.
Recap the cuvette.
5. Allow sample to react for any required wait time.*
6. If additional tablet or reagent is required, add,
crush, and stir with stir rod. Allow sample to react
for any required wait time.
7. Wipe any moisture or fingerprints off of cuvette
and place in colorimeter.
8. Press
zero•test
key.
9. If reading is correct press,
power
key to
exit calibration.
10. If adjustment is required, press
mode
key to
increase the value or press
zero•test
key to
decrease value.
11. Press
power
key twice to accept calibration.
: :
will be displayed briefly.
* NOTE: When calibrating Zinc and Copper, a
reading must be taken at this point. Proceed to
step 7. Perform reading, then add additional
reagent in step 6 and complete calibration.
After troubleshooting all components of your measurement system, contact The
Technical Edge
SM
for Orion products. Within the United States call 1.800.225.1480,
outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the
Middle East and Africa, contact your local authorized dealer. For the most current
contact information, visit www.thermo.com.
235625-001 Rev. C
Analyze • Detect • Measure • Control™
Calibration
Orion AQUAfast
®
II